Manage QC Data Run
Page Content
Manage QC Data Run / Enter Comments
Managed
- Select this option to add a comment for a data point and INCLUDE it in QC calculations.
Not Managed
- Select this option to add a comment for a data point and EXCLUDE it from QC calculations.
Managed
Not Managed
Select Comment Type:
No action - ACCEPT RUN
Repeat QC same vial - results in range - ACCEPT RUN
Repeat QC same vial- results out of range - REJECT RUN
Analyzed new vial - results in range - ACCEPT RUN
QC run in wrong file - REJECT RUN
QC vial QNS, repeat with new QC vial -REJECT RUN
Maintenance performed, repeat QC same vial - ACCEPT RUN
Analyzer serviced, repeat QC same vial - ACCEPT RUN
Instrument issues: Unable to perform testing. No patient results reported. - REJECT RUN
Calibration Verification - repeat QC - ACCEPT RUN
Reagent QNS. Replace reagent and repeat testing - REJECT RUN
WARNING – one 2SD failure – ACCEPT RUN
FAILURE – one 3SD failure – REJECT RUN
New lot number of QC, mean not established - ACCEPT RUN
Enter Comment Description: